FDA Recall Terminated

BleaseFocus Anesthesia Workstation and 700/900 Series Ventilators are anesthesia gas machine. In most cases, the customer ordered both a BleaseFocus anesthesia workstation and a 700/900 ventilator. In some cases, the customer may order a 700/900 stand-alone ventilator would could be used with a BleaseFocus anesthesia workstation at a later date.

Recall: Z-0565-2013 · Initiated November 2, 2012

Recall

Recall Number
Z-0565-2013
Event Number
63714
Firm
Del Mar Reynolds Medical, Ltd. 1-2 Harforde Ct., Foxholes Business Park Hertford United Kingdom
FEI Number
3002807992
Product Code
BSZ
Status
Terminated
Root Cause
Labeling design
Initiated
November 2, 2012
Posted
December 19, 2012
Terminated
July 11, 2013

Description

BleaseFocus Anesthesia Workstation and 700/900 Series Ventilators are anesthesia gas machine. In most cases, the customer ordered both a BleaseFocus anesthesia workstation and a 700/900 ventilator. In some cases, the customer may order a 700/900 stand-alone ventilator would could be used with a BleaseFocus anesthesia workstation at a later date.

Reason

The BleaseFocus Anesthesia Workstation and 700/900 Series Ventilators were shipped with manuals that describe a secondary O2 Supply Failure alarm which has not been incorporated into the system. The manuals were prematurely updated for a feature that is not yet available.

Action

Spacelabs has sent the Urgent -Medical Device Correction-Spacelabs Healthcare BleaseFocus Anesthesia Workstation Blease 700/900 Series Ventilators letter, dated 16 November 2012, to US customers. On November 26, 2012, a customer letter (translated as necessary) was emailed to all international subsidiaries and distributors of record. Customers are advised that the primary alarm will sound only in the event of an O2 supply failure and they will be provided with the corrected user and service manuals that do not reference a secondary O2 supply failure alarm. Customers with questions can call Spacelabs Healthcare at 1-800-522-7025 and select 2 for Technical Support.

Distribution

Worldwide Distribution--USA (nationwide) including states of North Carolina and Oklahoma and the countries of Algeria, Argentina, Bahrain, Bangladesh, Bolivia, Chile, China, Columbia, Dominican Republic, Ecuador, Egypt, Finland, Ghana, Great Britian, Guatemala, Honduras, Hong Kong, India, Indonesia, Iraq, Jamaica, Jordan, Lebanon, Libya, Malaysia, Mauritius, Mexico, Moldova, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Pakistan, Palestine, Paraguay, Peru, Philippines, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Sudan, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates , Uruguay, Venezuela, and Vietnam.

Quantity

3 in US and 3121 Outside US