FDA Recall Terminated

Stille Sonesta Gynecology and Urology Table 6302 and 6302U; an electric cystometric table; Stille AB, Solna, Sweden. Used to position the subject/patient in a desired position during urodynamic, general examinations, cystoscopy, ultrasound, biofeedback, vasectomy and OBGYN procedures.

Recall: Z-0565-2010 · Initiated November 3, 2009

Recall

Recall Number
Z-0565-2010
Event Number
53817
Firm
Stille AB Gardsvagen
FEI Number
3003696170
Product Code
MMZ
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
November 3, 2009
Posted
December 23, 2009
Terminated
December 10, 2010
Address
14 Solna Sweden

Description

Stille Sonesta Gynecology and Urology Table 6302 and 6302U; an electric cystometric table; Stille AB, Solna, Sweden. Used to position the subject/patient in a desired position during urodynamic, general examinations, cystoscopy, ultrasound, biofeedback, vasectomy and OBGYN procedures.

Reason

There is a potential hazard to the patient if the bolts holding the side rail break, causing the accessories attached at the side rails of the tables and chairs to fall off during a medical procedure.

Action

Stille sent Field Safety Notices dated 10/30/09 to the direct accounts, informing them that there is a potential hazard with the Sonesta tables if the bolts holding the side rail break, causing the accessories attached at the side rails of the tables and chairs to fall off during a medical procedure. Stille developed a retrofit kit consisting of six side rail bolts and instructions for installing the bolts, which were sent to the end users with the safety notice. The accounts were requested to replace the side rail bolts with the new bolts provided, and to complete and return the enclosed acknowledgement card to Stille by fax or e-mail as soon as possible and no later than 12/11/09. Contact Stille Surgical Inc. at 224-612-5404 for assistance.

Distribution

Worldwide Distribution -- Including countries of USA and Canada.

Quantity

68 units total