FDA Recall
Terminated
CELL-DYN Diluent/Sheath Reagent, for use with CELL-DYN Ruby, Sapphire and 4000 systems. List #01H73-01. Packaged in 20 Liter cubitainers.
Recall: Z-0565-2007
·
Initiated December 22, 2006
Recall
- Recall Number
- Z-0565-2007
- Event Number
- 37234
- Firm
- Abbott Laboratories
- FEI Number
- 2919069
- Product Code
- GIF
- Status
- Terminated
- Root Cause
- Other
- Initiated
- December 22, 2006
- Posted
- March 1, 2007
- Terminated
- November 28, 2007
- Address
- 5440 Patrick Henry Dr, Santa Clara, CA, 95054-1113
Description
CELL-DYN Diluent/Sheath Reagent, for use with CELL-DYN Ruby, Sapphire and 4000 systems. List #01H73-01. Packaged in 20 Liter cubitainers.
Reason
Results for tests of RBC, MCV, RDW, MPV and PLT may be falsely elevated or falsely decreased when the Cell Dyn Diluent/Sheath reagent is used with Cell-Dyn 4000, Ruby and Sapphire systems.
Action
Recall initated 12/22/2006. Product letters were sent to consignees, as well as a customer reply form. The letters were sent with customer reply forms. Responses will be tracked via the response forms included with letters.
Distribution
Nationwide, including California and Nevada. An Abbott Laboratories facility in Hong Kong.
Quantity
652 (20 L. each) cubitainers