FDA Recall Terminated

CIVCO Needle Guide Bracket, REF P03122-06, Reusable non-sterile bracket for use with SonoSite L25/L25e transducers, Mfg by: CIVCO, Kalona, IA.

Recall: Z-0564-2019 · Initiated November 2, 2018

Recall

Recall Number
Z-0564-2019
Event Number
81361
Firm
CIVCO Medical Instruments Co., Inc.
FEI Number
3012763756
Product Code
ITX
Status
Terminated
Root Cause
Other
Initiated
November 2, 2018
Terminated
November 20, 2020
Address
2301 Jones Blvd, Coralville, IA, 52241-3469

Description

CIVCO Needle Guide Bracket, REF P03122-06, Reusable non-sterile bracket for use with SonoSite L25/L25e transducers, Mfg by: CIVCO, Kalona, IA.

Reason

The needle guide bracket may puncture the sterile cover applied over the transducer and bracket during assembly.

Action

The recalling firm began issuing letters dated 10/30/2918 via email on 11/2/2018 requesting return of the boxes of product or reusable brackets that may be in their inventory.

Distribution

Foreign distribution was made to Canada, Argentina, Australia, Austria, Belgium, Brazil, Chile, China, Colombia, Croatia, Cyprus, Denmark, Estonia, France, Germany, Hong Kong, India, Israel, Italy, Japan, Latvia, Lebanon, Lithuania, Netherlands, New Zealand, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Taiwan, Trinidad/Tobago, Turkey, Ukraine, United Arab Emirates, and the United Kingdom.

Quantity

948 boxes