FDA Recall Open, Classified

SAFE-T-FILL Plastic Blood Gas Capillary Tubes, 115 L; Prepared with Balanced Heparin Model/Catalog Number: 06 0185 Product Description: A clear, plastic capillary blood collection tube

Recall: Z-0563-2026 · Initiated September 26, 2025

Recall

Recall Number
Z-0563-2026
Event Number
97592
Firm
ASP Global, LLC. dba Anatomy Supply Partners, LLC.
FEI Number
3004824601
Product Code
GIO
Status
Open, Classified
Root Cause
Process change control
Initiated
September 26, 2025
Posted
November 19, 2025
Address
7800 Third Flag Pkwy, Austell, GA, 30168-7657

Description

SAFE-T-FILL Plastic Blood Gas Capillary Tubes, 115 L; Prepared with Balanced Heparin Model/Catalog Number: 06 0185 Product Description: A clear, plastic capillary blood collection tube

Reason

Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.

Action

On 09/26-29/2025, ASP Global emailed an URGENT MEDICAL DEVICE RECALL Letter to customers. ASP Global, LLC, is voluntarily recalling all RAM Scientific SAFE-T-FILL products within remaining expiry. ASP Global was made aware that our contract manufacturer for the RAM Scientific SAFE-T-FILL products was issued an FDA Warning Letter citing significant violations of the Quality System requirements, including inadequate production and process controls, as well as design controls. Customers are instructed to: a) Discontinue use of the recalled products immediately. b) Quarantine any remaining inventory to prevent further use. c) If critical or abnormal test results were obtained using the affected SAFE-T-FILL products, ensure they are verified through repeat or confirmatory testing in accordance with institutional protocols (e.g., venous confirmation of abnormal pediatric screening results). d) Dispose remaining inventory in accordance with your local facility guidelines or notify ASP Global at [email protected] for return authorization and disposal. e) Complete and return the Recall Response Form to confirm receipt and actions to [email protected]. For questions or assistance - contact [email protected] (855)-867-2190| Mon Fri, 9:00 AM 5:00 PM ET

Distribution

Distribution US nationwide and Canada.

Quantity

121750