CoverStainer Slide Rack The CoverStainer Slide Rack is an accessory to the CoverStainer instrument. The purpose is to hold microscopic slides, during transfer through the staining process. The rack can hold up to 10 slides at a time, and is necessary for optimal CoverStainer instrument performance.
Recall
- Recall Number
- Z-0563-2013
- Event Number
- 63807
- Firm
- Dako Denmark A/S Produktionsvej 42 Glostrup Denmark
- FEI Number
- 3002806778
- Product Code
- KPA
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- November 26, 2012
- Posted
- December 19, 2012
- Terminated
- May 14, 2013
Description
CoverStainer Slide Rack The CoverStainer Slide Rack is an accessory to the CoverStainer instrument. The purpose is to hold microscopic slides, during transfer through the staining process. The rack can hold up to 10 slides at a time, and is necessary for optimal CoverStainer instrument performance.
The expected life time of the CoverStainer slide racks (CS 10330) does not meet established performance requirements, originating from variations in the manufacturing process.
Dako sent a Voluntary Product Recall Notification dated November 13, 2012, to all affected customers. This voluntary recall notification was initiated because the expected life time of the CoverStainer slide racks (CS 10330) does not meet established performance requirements, originating from variations in the manufacturing process. The result can be that slides are falling out of the rack during the staining process. While this has been observed at only a few customer sites, the corrective action includes all CoverStainer installations. Customers were notified that Dako would correct the rack situation by using a new manufacturer and has imroved the rack design. Customers with questions were instructed to contact their local Dako representative. For questions regarding this recall call 805-566-3036.
Worldwide Distribution - USA (nationwide) and the following countries: Australia France Rep. of Korea Belgium Germany Russia Brazil Israel Saudi Arabia Canada Italy Spain Chile Japan Sweden China Mexico Taiwan Denmark Netherlands Turkey Ecuador Norway UK Finland Poland
25 Units US (50-100 racks per unit) 106 total