FDA Recall Terminated

Centrifuge Model CF-6 with lid lock. (Block Model number OCT-VI). Units are distributed to Bloch Scientific, Holbrook, NY for Bloch''s customers MFT and Cascade.

Recall: Z-0563-06 · Initiated February 4, 2005

Recall

Recall Number
Z-0563-06
Event Number
34385
Firm
Cygnus LLC
FEI Number
1000121662
Product Code
JQC
Status
Terminated
Root Cause
Other
Initiated
February 4, 2005
Posted
February 28, 2006
Terminated
February 1, 2007
Address
510 E 41st St, Paterson, NJ, 07504-1210

Description

Centrifuge Model CF-6 with lid lock. (Block Model number OCT-VI). Units are distributed to Bloch Scientific, Holbrook, NY for Bloch''s customers MFT and Cascade.

Reason

The units were exhibiting tolerance stacking in the electronic circuit timer that controls the electric motor that spins the centrifuge and applies the centrifuge brake.

Action

A recall notificaiton was sent to the firm''s customers by e-mail on 2/14/2005. Distributor conducted sub-recall.

Distribution

All product was shipped domestically to the firm''s distributor, Bock Scientific, 1170 Lincoln Avenue, Unit 11, Holbrook, NY 11741

Quantity

20 units