Dimension(R) Chemistry Wash (RD701), Siemens Material Number (SMN) 10445052 Product Usage: Used in the Dimension(R) system with the heterogeneous module
Recall
- Recall Number
- Z-0561-2014
- Event Number
- 65131
- Firm
- Siemens Healthcare Diagnostics, Inc.
- FEI Number
- 2517506
- Product Code
- JMG
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- May 6, 2013
- Posted
- December 24, 2013
- Terminated
- August 27, 2014
- Address
- 500 Gbc Dr PO BOX 6101, Ms 514, Newark, DE, 19702-2466
Description
Dimension(R) Chemistry Wash (RD701), Siemens Material Number (SMN) 10445052 Product Usage: Used in the Dimension(R) system with the heterogeneous module
Siemens Healthcare Diagnostics received complaints of Chemistry Wash causing QC and patient sample result shifts on CTN1, LTN1, FT4, TSH, PBNP and LPBN when using Chemistry Wash lots RD23031, RD23111, RD23241, RD23311, RD23391, and RD23461. These lots shipped between 12/12/2012 and 3/22/2013. Internal investigation has confirmed shifts can be encountered in two scenarios: 1 - When an affected l
Siemens sent an Urgent Medical Device Recall letter dated May 2013 to all affected customers. The letter identified the product and lots, the reason for the field action, the risk to health and the action(s) to be taken. Customers were instructed to discontinue use and discard inventory of the affected product. Siemens will replace any unused inventory of the affected lots at no charge. For questions contact your Siemens Technical Center or your local Siemens technical support representative.
US Nationwide and OUS.
17,719