Xoran MiniCAT System for ENT (server and work station), Xoran Technologies, Ann Arbor, MI. Cone beam CT for sinus, temporal bones, and maxillofacial imaging.

Recall: Z-0561-2010 · Initiated September 17, 2008

Recall

Recall Number
Z-0561-2010
Event Number
47370
Firm
Xoran Technologies, Inc.
FEI Number
3004198450
Product Code
JAK
Status
Terminated
Root Cause
Device Design
Initiated
September 17, 2008
Posted
December 23, 2009
Terminated
December 23, 2009
Address
309 N 1st St, Ann Arbor, MI, 48103-3301

Description

Xoran MiniCAT System for ENT (server and work station), Xoran Technologies, Ann Arbor, MI. Cone beam CT for sinus, temporal bones, and maxillofacial imaging.

Reason

The system may overheat, causing poor image quality, interruption in image acquisition and/or damage to the system. Also, if the system is not allowed to cycle down prior to being restarted after an emergency stop, the x-ray and gantry motions will both resume.

Action

A letter dated 9/17/08 was sent to each account telling them that the PLC would be upgraded. The firm visited each account and made corrections.

Distribution

Nationwide Distribution

Quantity

82