FDA Recall
Terminated
Xoran MiniCAT System for ENT (server and work station), Xoran Technologies, Ann Arbor, MI. Cone beam CT for sinus, temporal bones, and maxillofacial imaging.
Recall: Z-0561-2010
·
Initiated September 17, 2008
Recall
- Recall Number
- Z-0561-2010
- Event Number
- 47370
- Firm
- Xoran Technologies, Inc.
- FEI Number
- 3004198450
- Product Code
- JAK
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- September 17, 2008
- Posted
- December 23, 2009
- Terminated
- December 23, 2009
- Address
- 309 N 1st St, Ann Arbor, MI, 48103-3301
Description
Xoran MiniCAT System for ENT (server and work station), Xoran Technologies, Ann Arbor, MI. Cone beam CT for sinus, temporal bones, and maxillofacial imaging.
Reason
The system may overheat, causing poor image quality, interruption in image acquisition and/or damage to the system. Also, if the system is not allowed to cycle down prior to being restarted after an emergency stop, the x-ray and gantry motions will both resume.
Action
A letter dated 9/17/08 was sent to each account telling them that the PLC would be upgraded. The firm visited each account and made corrections.
Distribution
Nationwide Distribution
Quantity
82