SODASORB Carbon Dioxide Absorbent, Medical Grade (16%) Soda Lime USP-NF with an ethyl violet indicator which turns from white to violet as absorbent is depleted; Manufactured by: W.R. Grace & Co., 6050 West 51st St., Chicago, IL 60638; the product is packaged in pre-paks pre-filled 2.5 lb. canisters, 12 per carton The indicator soda lime product is used for absorption of carbon dioxide in anesthesia rebreathing systems and respiratory therapy equipment with oxygen, nitrous oxide, halothane, enflurane, sevoflurane, desflurane and isoflurane.

Recall: Z-0560-2011 · Initiated October 20, 2010

Recall

Recall Number
Z-0560-2011
Event Number
57064
Firm
Darex Container Products Div of W.R. Grace & Co.
FEI Number
3003940942
Product Code
CBL
Status
Terminated
Root Cause
Packaging process control
Initiated
October 20, 2010
Posted
December 8, 2010
Terminated
September 21, 2011
Address
6050 W 51st St, Chicago, IL, 60638-1405

Description

SODASORB Carbon Dioxide Absorbent, Medical Grade (16%) Soda Lime USP-NF with an ethyl violet indicator which turns from white to violet as absorbent is depleted; Manufactured by: W.R. Grace & Co., 6050 West 51st St., Chicago, IL 60638; the product is packaged in pre-paks pre-filled 2.5 lb. canisters, 12 per carton The indicator soda lime product is used for absorption of carbon dioxide in anesthesia rebreathing systems and respiratory therapy equipment with oxygen, nitrous oxide, halothane, enflurane, sevoflurane, desflurane and isoflurane.

Reason

Pre-Pak units of SODASORB CO2 absorbent manufactured prior to August 2010 may have missing or illegible expiration dates and/or batch numbers.

Action

Darex/W.R. Grace sent a Customer Notification to their sole direct account of the recall and the account provided Darex/W.R. Grace with lists of their customers who received the recalled products. Darex then sent undated recall letters to those accounts on 10/20-22/10. The accounts were informed of the product, the problem, and the action that should be taken. The accounts were warned not to use Pre-Pak units with missing or illegible expiration dates and/or batch numbers and were requested to examine their inventories of Pre-Pak units, placing all units that had missing or illegible expiration dates and/or batch numbers on hold for destruction and replacement. The accounts were instructed to call W.R. Grace at 1-800-492-7632 to arrange for replacement. The accounts were also requested to complete and return the reply postcard, indicating how much product was destroyed.

Distribution

Nationwide Distribution

Quantity

95,542 cartons