Evolut (TM) R System. With the following model numbers: a. EVOLUTR-23, Labeled as VLV EVOLUTR-23 VT IR 6L COMM OUS, VLV EVOLUTR-23 TAV EVOLUT R CE 29L MX, VLV EVOLUTR-23 TAV CE AU 2Y MX, VLV EVOLUTR-23 TAV JPN MX, VLV EVOLUTR-23 GREY 29L COMM OUS, VLV EVOLUTR-23 TAV EVOLUT R CE 7L MX, VLV EVOLUTR-23 TAV EVOLUT R CE 26L MX, VLV EVOLUTR-23 TAV CA 2Y MX, VLV EVOLUTR-23 GREY 6L COMM OUS, VLV EVOLUTR-23-US TAV 23 US COMM MX; b. EVOLUTR-23-C, Labeled as VLV EVOLUTR-23-C TAV EVOLUT R US IDE MEX; c. EVOLUTR-23-US, Labeled as VLV EVOLUTR-23-US GREY COMM US, VLV EVOLUTR-23-US TAV 23 US COMM MX; d. EVOLUTR-26, Labeled as VLV EVOLUTR-26 TAV EVOLUT R CE 29L MX VLV EVOLUTR-26 VT IR 6L COMM OUS, VLV EVOLUTR-26 TAV EVOLUT R CE 7L MX, LV EVOLUTR-26 TAV CE AU 2Y MX, VLV EVOLUTR-26 TAV JPN MX, VLV EVOLUTR-26 GREY 29L COMM OUS, VLV EVOLUTR-26-US TAV 26 US COMM MX, VLV EVOLUTR-26 TAV CLN OUS MX, VLV EVOLUTR-26 TAV CA 2Y MX, VLV EVOLUTR-26 TAV CE CA 2Y MX, VLV EVOLUTR-26 GREY 6L COMM OUS; e. EVOLUTR-26-C, Labeled as VLV EVOLUTR-26-C TAV EVOLUT R US IDE MEX, VLV EVOLUTR-26-C TAV CLN US MX; f. EVOLUTR-26-US, Labeled as VLV EVOLUTR-26 TAV EVOLUT R CE 29L MX, VLV EVOLUTR-26-US GREY COMM US, VLV EVOLUTR-26-US TAV 26 US COMM MX; g. EVOLUTR-29, Labeled as VLV EVOLUTR-29 TAV EVOLUT R CE 29L MX, VLV EVOLUTR-29 GREY 6L COMM OUS, VLV EVOLUTR-29 VT IR 6L COMM OUS, VLV EVOLUTR-29 TAV CE AU 2Y MX, VLV EVOLUTR-29 TAV JPN MX, VLV EVOLUTR-29 GREY 29L COMM OUS, VLV EVOLUTR-29 TAV EVOLUT R CE 7L MX, VLV EVOLUTR-29 TAV CLN OUS MX, VLV EVOLUTR-29 TAV CA 2Y MX, VLV EVOLUTR-29 TAV CE CA 2Y MX, VLV EVOLUTR-29-US TAV 29 US COMM MX; h. EVOLUTR-29-C, Labeled as VLV EVOLUTR-29-C TAV EVOLUT R US IDE MEX, LV EVOLUTR-29-C TAV CLN US MX, VLV EVOLUTR-29-US TAV 29 US COMM MX; i. EVOLUTR-29-US, Labeled as VLV EVOLUTR-29-US GREY COMM US, VLV EVOLUTR-29-US TAV 29 US COMM MX, VLV EVOLUTR-29 TAV EVOLUT R CE 29L MX; j. EVOLUTR-34, Labeled as VLV EVOLUTR-34 BLUE 29L COMM OUS, VLV EVOLUTR-34 VT IR 6L COMM OUS, VLV EVOLUTR-34 BLUE 6L COMM OUS, VLV EVOLUTR-34 TAV CE AU 2Y MX, VLV EVOLUTR-34 TAV JPN MX, VLV EVOLUTR-34 TAV CLN OUS MX, VLV EVOLUTR-34 TAV CA 2Y MX, VLV EVOLUTR-34 TAV CE CA 2Y MX, VLV EVOLUTR-34 TAV EVOLUT R CE 29L 1Y MX, VLV EVOLUTR-34 TAV EVO R CE 29L 2Y MX SA, DRAPE 9732722 TUBE STERILE O-ARM 20PK; k. EVOLUTR-34-C, Labeled as VLV EVOLUTR-34-C TAV EVOLUT R US IDE MX, VLV EVOLUTR-34-C TAV CLN US MX; l. EVOLUTR-34-US, Labeled as VLV EVOLUTR-34-US BLUE COMM US, VLV EVOLUTR-34-US TAV 34 US COMM MX, VLV EVOLUTR-34 TAV EVOLUT R CE 29L 1Y MX, VLV EVOLUTR-34-US TAV 34 US COMM MX SA.

Recall: Z-0559-2021 · Initiated October 23, 2020

Recall

Recall Number
Z-0559-2021
Event Number
86681
Firm
Medtronic Heart Valves Division
FEI Number
2025587
Product Code
NPT
Status
Terminated
Root Cause
Labeling design
Initiated
October 23, 2020
Terminated
November 22, 2022
Address
1851 E Deere Ave, Santa Ana, CA, 92705-5720

Description

Evolut (TM) R System. With the following model numbers: a. EVOLUTR-23, Labeled as VLV EVOLUTR-23 VT IR 6L COMM OUS, VLV EVOLUTR-23 TAV EVOLUT R CE 29L MX, VLV EVOLUTR-23 TAV CE AU 2Y MX, VLV EVOLUTR-23 TAV JPN MX, VLV EVOLUTR-23 GREY 29L COMM OUS, VLV EVOLUTR-23 TAV EVOLUT R CE 7L MX, VLV EVOLUTR-23 TAV EVOLUT R CE 26L MX, VLV EVOLUTR-23 TAV CA 2Y MX, VLV EVOLUTR-23 GREY 6L COMM OUS, VLV EVOLUTR-23-US TAV 23 US COMM MX; b. EVOLUTR-23-C, Labeled as VLV EVOLUTR-23-C TAV EVOLUT R US IDE MEX; c. EVOLUTR-23-US, Labeled as VLV EVOLUTR-23-US GREY COMM US, VLV EVOLUTR-23-US TAV 23 US COMM MX; d. EVOLUTR-26, Labeled as VLV EVOLUTR-26 TAV EVOLUT R CE 29L MX VLV EVOLUTR-26 VT IR 6L COMM OUS, VLV EVOLUTR-26 TAV EVOLUT R CE 7L MX, LV EVOLUTR-26 TAV CE AU 2Y MX, VLV EVOLUTR-26 TAV JPN MX, VLV EVOLUTR-26 GREY 29L COMM OUS, VLV EVOLUTR-26-US TAV 26 US COMM MX, VLV EVOLUTR-26 TAV CLN OUS MX, VLV EVOLUTR-26 TAV CA 2Y MX, VLV EVOLUTR-26 TAV CE CA 2Y MX, VLV EVOLUTR-26 GREY 6L COMM OUS; e. EVOLUTR-26-C, Labeled as VLV EVOLUTR-26-C TAV EVOLUT R US IDE MEX, VLV EVOLUTR-26-C TAV CLN US MX; f. EVOLUTR-26-US, Labeled as VLV EVOLUTR-26 TAV EVOLUT R CE 29L MX, VLV EVOLUTR-26-US GREY COMM US, VLV EVOLUTR-26-US TAV 26 US COMM MX; g. EVOLUTR-29, Labeled as VLV EVOLUTR-29 TAV EVOLUT R CE 29L MX, VLV EVOLUTR-29 GREY 6L COMM OUS, VLV EVOLUTR-29 VT IR 6L COMM OUS, VLV EVOLUTR-29 TAV CE AU 2Y MX, VLV EVOLUTR-29 TAV JPN MX, VLV EVOLUTR-29 GREY 29L COMM OUS, VLV EVOLUTR-29 TAV EVOLUT R CE 7L MX, VLV EVOLUTR-29 TAV CLN OUS MX, VLV EVOLUTR-29 TAV CA 2Y MX, VLV EVOLUTR-29 TAV CE CA 2Y MX, VLV EVOLUTR-29-US TAV 29 US COMM MX; h. EVOLUTR-29-C, Labeled as VLV EVOLUTR-29-C TAV EVOLUT R US IDE MEX, LV EVOLUTR-29-C TAV CLN US MX, VLV EVOLUTR-29-US TAV 29 US COMM MX; i. EVOLUTR-29-US, Labeled as VLV EVOLUTR-29-US GREY COMM US, VLV EVOLUTR-29-US TAV 29 US COMM MX, VLV EVOLUTR-29 TAV EVOLUT R CE 29L MX; j. EVOLUTR-34, Labeled as VLV EVOLUTR-34 BLUE 29L COMM OUS, VLV EVOLUTR-34 VT IR 6L COMM OUS, VLV EVOLUTR-34 BLUE 6L COMM OUS, VLV EVOLUTR-34 TAV CE AU 2Y MX, VLV EVOLUTR-34 TAV JPN MX, VLV EVOLUTR-34 TAV CLN OUS MX, VLV EVOLUTR-34 TAV CA 2Y MX, VLV EVOLUTR-34 TAV CE CA 2Y MX, VLV EVOLUTR-34 TAV EVOLUT R CE 29L 1Y MX, VLV EVOLUTR-34 TAV EVO R CE 29L 2Y MX SA, DRAPE 9732722 TUBE STERILE O-ARM 20PK; k. EVOLUTR-34-C, Labeled as VLV EVOLUTR-34-C TAV EVOLUT R US IDE MX, VLV EVOLUTR-34-C TAV CLN US MX; l. EVOLUTR-34-US, Labeled as VLV EVOLUTR-34-US BLUE COMM US, VLV EVOLUTR-34-US TAV 34 US COMM MX, VLV EVOLUTR-34 TAV EVOLUT R CE 29L 1Y MX, VLV EVOLUTR-34-US TAV 34 US COMM MX SA.

Reason

Potential for Transcatheter Aortic Valve (TAV) leaflet damage when performing a post-implant balloon dilatation (PID).

Action

On 10/23/2020, the firm sent an "Urgent Medical Device Correction" Notification Letter to customers via UPS 2-day delivery informing them that the Recalling Firm has received complaint of valve leaflet damage occurring following a post-implant balloon dilatation (PID) resulting in moderate or severe aortic insufficiency which were detected acutely or during follow up. The Recalling Firm has identified the potential cause of the damage which is due to over-expansion of the narrowest portion (waist) of the Transcatheter Aortic Valves (TAVs). The Recalling Firm is also stating that depending on the choice of balloon, the physicians must consider two factors that may lead to over-expansion of the waist of the TAV: 1.The pressure the balloon is inflated to when performing PID 2.Balloon size used for PID The Recalling Firm is providing detailed guidance on considering the above two facts and is providing in an Appendix (Appendix A) with the Customer Notification Letter. In addition, the Appendix A is providing updated precautionary instructions related to PID. The Recalling Firm is also informing its customers that for patients who have been, or will be, treated with the affected product, should continue to be managed according to their standard patient management protocols. Physicians are also asked to: - Review the updated instructions provided in Appendix A. - Share this information with other physicians in your facility who use the affected product. - After review of this information, complete the enclosed Physician Confirmation Certificate and email to [email protected]. Any Questions can be directed to the Recalling Firm's Field Representatives or Lifeline Technical Services at 1-877-526-7890, Monday-Friday, 8am-5pm Central Time.

Distribution

Worldwide distribution - U.S. Nationwide distribution and the countries of Panama, Denmark, Austria, Germany, Finland, Sweden, Colombia, India, Mexico, Brazil, Spain, United Kingdom, Australia, Ireland, Italy, Argentina, Japan, Poland, Russian Federation, Hungary, Kazakhstan, Luxembourg, Portugal, Norway, Egypt, Saudi Arabia.

Quantity

172,851 units