VITROS Anti-HBs Calibrators & VITROS Anti-HBs Reagent Packs
Recall
- Recall Number
- Z-0557-2014
- Event Number
- 66900
- Firm
- Ortho-Clinical Diagnostics
- FEI Number
- 1000136573
- Product Code
- LOM
- Status
- Terminated
- Root Cause
- Other
- Initiated
- October 25, 2013
- Posted
- December 20, 2013
- Terminated
- May 18, 2018
- Address
- 100 Indigo Creek Dr, Rochester, NY, 14626-5101
Description
VITROS Anti-HBs Calibrators & VITROS Anti-HBs Reagent Packs
Ortho Clinical Diagnostics issued a Product Correction Notification for VITROS Immunodiagnostic Products Anti-HBs Calibrators and Anti-HBs Quantitative Reagent Packs due to false positive results.
On October 25, 2013, Ortho Clinical Diagnostics issued product correction notifications to their US customers via US Postal Service Priority Mail and email notices to their foreign affiliates informing them of the recall. Ortho Clinical Diagnostics advised their customers to review previously reported results between > 12 and <30 mIU/mL and consult the Laboratory Medical Director. Customers were instructed to utilize alternative methods for anti-HBs testing. Customers are advised to seek alternative methods for anti-HBs testing. For additional information, please contact the OCD Customer Technical Service representative at 1-800-421-3311.
Nationwide distribution - all states.
1,920 kits