FDA Recall
Open, Classified
Non-ablative Fractional Laser Systems, Model: WFB-01
Recall: Z-0555-2025
·
Initiated August 23, 2024
Recall
- Recall Number
- Z-0555-2025
- Event Number
- 95831
- Firm
- WINGDERM ELECTRO-OPTICS LTD. Room 312 Building D-3, Dongsheng Science Park, No.66 Of Xixiaokou Road, Haidian District Beijing China
- FEI Number
- 3014310779
- Product Code
- ONG
- Status
- Open, Classified
- Root Cause
- Radiation Control for Health and Safety Act
- Initiated
- August 23, 2024
- Posted
- December 5, 2024
Description
Non-ablative Fractional Laser Systems, Model: WFB-01
Reason
Model WFB-01 sold to US customers before Aug 23, 2022 has been found to have a software bug (6 units). All WFB-01 models sold to US customers found to have labeling non-compliances (22 units)
Action
The firm sent a notification letter to the relevant distribution on June 6, 2024. Products sold before August 23, 2022, will require a software version upgrade to F1-B-1.1.012 or Higher.
Distribution
Worldwide Distribution