FDA Recall
Terminated
TD60 Transmitter battery charger used with the BeneVision Central Station, Part number 115-030108-00
Recall: Z-0554-2018
·
Initiated July 14, 2017
Recall
- Recall Number
- Z-0554-2018
- Event Number
- 79005
- Firm
- Mindray DS USA, Inc. dba Mindray North America
- FEI Number
- 2221819
- Product Code
- MSX
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- July 14, 2017
- Terminated
- December 17, 2018
- Address
- 800 Macarthur Blvd, Mahwah, NJ, 07430-2001
Description
TD60 Transmitter battery charger used with the BeneVision Central Station, Part number 115-030108-00
Reason
A damaged circuit board in the battery charger can cause the battery to overcharge and the battery case to swell and separate.
Action
Certified letters were sent to all affected customers on 07/14/2017, to offer a replacement of the battery charger.
Distribution
Distributed in 26 states: AL, CA, CO, FL, GA, IA, IN, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NM, NV, NY, OH, PA, SD, TX, UT, WA, and WY; and Puerto Rico.
Quantity
203