Cmax Surgical Table, Model No. 2182625, Manufactured by STERIS Corporation, 2720 Gunter Park East, Montgomery, AL 36109 USA, TEL: 334 277 6660, FAX: 334 271 5450

Recall: Z-0554-2007 · Initiated February 6, 2007

Recall

Recall Number
Z-0554-2007
Event Number
37349
Firm
Steris Corporation
FEI Number
1043572
Product Code
GDC
Status
Terminated
Root Cause
Other
Initiated
February 6, 2007
Posted
February 23, 2007
Terminated
September 11, 2009
Address
2720 Gunter Park Drive East, Montgomery, AL, 36109-1410

Description

Cmax Surgical Table, Model No. 2182625, Manufactured by STERIS Corporation, 2720 Gunter Park East, Montgomery, AL 36109 USA, TEL: 334 277 6660, FAX: 334 271 5450

Reason

A variation in suppliers welding processes, at the inner section of the hydraulic lift column, of the surgical table, may result in the table top becoming unstable, which could result in injury to patients or staff.

Action

A certified letter was sent to all affected consignees on February 6, 2007 notifying them of the issue and informing them of the recall and the replacement of the affected tables with new tables.

Distribution

USA and Canada

Quantity

46 tables