FDA Recall Terminated

Medline Custom Sterile Surgical Packs which contain a Tyvek Breather Pouch, part number 32901, lot numbers 44134-01 and 43589-01; the breather pouch measures 25.25 inches x 30.75 inches, 5 mil thick; Medline Industries, Inc., Mundelein, IL 60060;

Recall: Z-0554-05 · Initiated March 4, 2005

Recall

Recall Number
Z-0554-05
Event Number
31315
Firm
Medline Industries, Inc.
FEI Number
3000717925
Product Code
LRO
Status
Terminated
Root Cause
Other
Initiated
March 4, 2005
Posted
March 11, 2005
Terminated
May 25, 2006
Address
1170 S Northpoint Blvd, Waukegan, IL, 60085-6757

Description

Medline Custom Sterile Surgical Packs which contain a Tyvek Breather Pouch, part number 32901, lot numbers 44134-01 and 43589-01; the breather pouch measures 25.25 inches x 30.75 inches, 5 mil thick; Medline Industries, Inc., Mundelein, IL 60060;

Reason

There is a potential for open side seals post-sterilization of the Tyvek breather pouches used in the custom sterile packs.

Action

Field Correction letters dated 3/4/05 were sent to thedirect account hospitals on the same date to the attention of the manager of surgery, informing them of the potential indadequate side seal of the Tyvek breather pouches in the packs which may cause a compromise in the sterility of the product. The accounts were provided a list of the affected pack sent to the account and were requested to inspect any stocks of the affected product in stock, using the enclosed diagrams as an aid in the inspection for the defective side seals. Any defective product found should be returned to Medline. Any questions were directed to Lara Simmons, QA/RA Director at 1-800-879-2781, ext. 6072.

Distribution

Nationwide

Quantity

31,497 cases