Symphony Staining System (High Volume Stainer), KPA, Model Number 2200000, Ventana Medical Systems Inc

Recall: Z-0553-2008 · Initiated October 30, 2007

Recall

Recall Number
Z-0553-2008
Event Number
45891
Firm
Ventana Medical Systems Inc
FEI Number
1000125183
Product Code
KEY
Status
Terminated
Root Cause
Component design/selection
Initiated
October 30, 2007
Posted
January 25, 2008
Terminated
April 14, 2008
Address
1910 E Innovation Park Dr, Tucson, AZ, 85755-1962

Description

Symphony Staining System (High Volume Stainer), KPA, Model Number 2200000, Ventana Medical Systems Inc

Reason

Fluid Leaks-The T fitting located in the Symphony Clear system of the Symphony High Volume Slide Stainer needs to be larger to securely hold the tubing in place to eliminate any possible fluid leak due to the tubing movement.

Action

The end user customers were verbally contacted starting in November 2007 to set up a scheduled visit to replace the fitting.

Distribution

Worldwide-USA, France and Japan

Quantity

55