Faaborg patient lift/hoist, battery operated lifts. The Solution/Nordic series lifts have electric V-shape base widening, with a weight capacity of 210 lbs. to 365 lbs. depending on the base model. Please note: the Faaborg Person Lift arm assembly can be interchanged on the various lift series bases. The arm assembly is labeled ''Faaborgliften Made in Denmark'', while the base plate reads in part ''Faaborg Rehab Technic Lift Model: VL-***** Serial No: **** Capacity: *** KG. Manufactured: **-**** 5600 Faaborg Phone+4570205553 Fax+4570205554'' A sticker is also present on the base reading ''Moving Solutions 800-228-7980 Fax 630-963-9588''. The Users Guide refers to the lift as a ''Faaborghoist''.

Recall: Z-0553-04 · Initiated January 21, 2004

Recall

Recall Number
Z-0553-04
Event Number
28199
Firm
Moving Solutions, Inc.
FEI Number
1000305914
Product Code
FSA
Status
Terminated
Root Cause
Other
Initiated
January 21, 2004
Posted
July 20, 2004
Terminated
October 22, 2007
Address
2550 Wisconsin Ave, Downers Grove, IL, 60515

Description

Faaborg patient lift/hoist, battery operated lifts. The Solution/Nordic series lifts have electric V-shape base widening, with a weight capacity of 210 lbs. to 365 lbs. depending on the base model. Please note: the Faaborg Person Lift arm assembly can be interchanged on the various lift series bases. The arm assembly is labeled ''Faaborgliften Made in Denmark'', while the base plate reads in part ''Faaborg Rehab Technic Lift Model: VL-***** Serial No: **** Capacity: *** KG. Manufactured: **-**** 5600 Faaborg Phone+4570205553 Fax+4570205554'' A sticker is also present on the base reading ''Moving Solutions 800-228-7980 Fax 630-963-9588''. The Users Guide refers to the lift as a ''Faaborghoist''.

Reason

Excessive wear of the hanger bar bolt, that connects the lift arm to the sling spreader bar, may cause the bolt to snap allowing the patient to fall.

Action

Letters were sent to all Faaborg lift customers on 11/29/01 and 01/24/04 advising the users that all bolts in connection with the hanger bar must be inspected, greased and tightened, and all defective or worn down bolts and washers must be changed. The 1/24/04 letters included an upgraded maintenance schedule and an Important Service/Safety Reminder, as well as a supply of hanger bar bolt washers. Any questions were directed to Moving Solutions' Service Department at 800-228-7980. Follow-up letters were sent to the accounts on 3/11/04, referencing the 1/21/04 letter and instructing the accounts not to use the lifts until the bolts have been inspected and retrofitted with the washers. Copies of the revised maintenance schedule and operating instructions were included.

Distribution

Nationwide to nursing homes, residential schools and home patients, and internationally to Canada.