FDA Recall Open, Classified

The System is designed as a modular system with components that include an Overhead Tube Suspension with tube/collimator, wallstand, Table, X-ray generator, and cleared wireless digital detectors. The list of detectors verified and validated for use with the Discovery XR656 HD system, including their specifications, are provided in the user documentation. The System generates diagnostic The components may be combined in different configurations to meet specific customer needs. In addition, upgrade configurations are available for predicate devices. The optional image pasting function enables the operator to stitch sequentially acquired radiographs into a single image. This 510(k) is to incorporate the VolumeRad advanced application that was currently available on the Discovery XR656 product onto the Discovery XR656 HD, as well as introduce a new Metal Artifact Reduction Algorithm, and an optional standalone console to take any HelixTM acquired images via DICOM (such as from a Discovery XR656 HD, Optima XR646 HD, or Optima XR240amx) and process the images independently of the system it was acquired on.

Recall: Z-0552-2024 · Initiated November 7, 2023

Recall

Recall Number
Z-0552-2024
Event Number
93479
Firm
GE Hangwei Medical Systems, Co. LTD West Area Bldg. 3, No.1 Yongchang North Road 1 Yongchangbeilu; Beijing Economic and Technological Development Area Beijing China
FEI Number
3002175190
Product Code
KPR
Status
Open, Classified
Root Cause
Process design
Initiated
November 7, 2023
Posted
December 14, 2023

Description

The System is designed as a modular system with components that include an Overhead Tube Suspension with tube/collimator, wallstand, Table, X-ray generator, and cleared wireless digital detectors. The list of detectors verified and validated for use with the Discovery XR656 HD system, including their specifications, are provided in the user documentation. The System generates diagnostic The components may be combined in different configurations to meet specific customer needs. In addition, upgrade configurations are available for predicate devices. The optional image pasting function enables the operator to stitch sequentially acquired radiographs into a single image. This 510(k) is to incorporate the VolumeRad advanced application that was currently available on the Discovery XR656 product onto the Discovery XR656 HD, as well as introduce a new Metal Artifact Reduction Algorithm, and an optional standalone console to take any HelixTM acquired images via DICOM (such as from a Discovery XR656 HD, Optima XR646 HD, or Optima XR240amx) and process the images independently of the system it was acquired on.

Reason

GE HealthCare has become aware of the potential that incorrect torque has been applied to certain bolts of the Definium Tempo and Definium Tempo Pro products which may result in a potential fall of elements of the Overhead Tube Suspension (OTS) assembly.

Action

GE HealthCare issued an URGENT MEDICAL DEVICE CORRECTION to its consignees on 11/07/2023 via FedEx. The notice explained the issue, hazard, and recommended the use of alternative device for patient exams until a service appointment can be scheduled to address the issue. Questions or concerns: Contact 1-800-467-1171 or local Service Representative

Distribution

US Nationwide distribution.

Quantity

134 units