Terumo Surflo Winged Infusion Set with Filter and Needle Protection (Surshield), Product Code: SV-S25FL35.
Recall
- Recall Number
- Z-0551-2016
- Event Number
- 72828
- Firm
- Terumo Medical Corp
- FEI Number
- 2243441
- Product Code
- MEG
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- November 23, 2015
- Posted
- December 24, 2015
- Terminated
- August 1, 2017
- Address
- 2101 Cottontail Lane, Somerset, NJ, 08873-1277
Description
Terumo Surflo Winged Infusion Set with Filter and Needle Protection (Surshield), Product Code: SV-S25FL35.
Potential improper bonding between the tube and luer for certain lots of Surflo Winged Infusion Set with Filter and Needle Protection (Surshield).
Terumo N.V. Europe notified their US Importer (Termumo Medical Corporation, Somerset, NJ) of the voluntary recall on 11/23/2015 via phone and email. Terumo Medical Corporation in turn notified their sole consignee on 11/24/2015 by email that included the manufacturers Recall Notification. The notification informed the customer of the reason for the recall; how to identify affected product; to immediately identify and quarantine any available stock; and complete and return the attached Reply Form.
California
80,500 units