FDA Recall Terminated

OXY-1 System, Model Number 001-0500-001. Used during cardiopulmonary bypass. OXY-1 System consists of Console, Pump Drive, E Drive, and Pole Mount. The disposable Pump Lung Unit PLU part number 001-0400-100 is sold separately and is not part of this recall.

Recall: Z-0550-2022 · Initiated December 14, 2021

Recall

Recall Number
Z-0550-2022
Event Number
89328
Firm
Abiomed, Inc.
FEI Number
1220648
Product Code
DTZ
Status
Terminated
Root Cause
Software design (manufacturing process)
Initiated
December 14, 2021
Posted
February 3, 2022
Terminated
February 28, 2023
Address
22 Cherry Hill Dr, Danvers, MA, 01923-2575

Description

OXY-1 System, Model Number 001-0500-001. Used during cardiopulmonary bypass. OXY-1 System consists of Console, Pump Drive, E Drive, and Pole Mount. The disposable Pump Lung Unit PLU part number 001-0400-100 is sold separately and is not part of this recall.

Reason

The firm has received complaints of the OXY-1 System Console experiencing a power interruption while in use, which disrupts therapy delivered to a patient on support. Disruption of therapy could lead to prolonged hypoxia.

Action

The firm began notifying customers by phone starting December 14, 2021. The firm followed up by sending Urgent Medical Device Recall letters to customers on January 20, 2022. The firm is requesting that customers return their devices. The firm does not recommend placing new patients on support. For patients currently on support with an affected system, the firm recommends physicians use their clinical judgement on whether it is appropriate to continue a patient on the OXY-1 System or to transition that patient to the next step in their clinical journey. The letter included instructions on how to return impacted product. If you have questions on returning product, If you have any questions, please contact Diane Welsh from Abiomed s customer service team at (978) 646-1467 or email [email protected]. The firm's Clinical field staff will be available to coordinate boxing and shipping of the OXY-1 system for return to Abiomed. If you have questions regarding this recall, please contact Ralph Barisano directly at (978) 882-8068.

Distribution

US distribution to FL, GA, IN, MD, NC, PA, TX.

Quantity

42 Units