FDA Recall Terminated

Canon Aquilion Prime SP, Multislice Helical CT Scanner, TSX-303B/8L - Product Usage: intended to acquire and display cross-sectional volumes of the whole body, including the head.

Recall: Z-0549-2021 · Initiated November 9, 2020

Recall

Recall Number
Z-0549-2021
Event Number
86864
Firm
Canon Medical System, USA, INC.
FEI Number
2020563
Product Code
JAK
Status
Terminated
Root Cause
Software design
Initiated
November 9, 2020
Terminated
December 16, 2022
Address
2441 Michelle Dr, Tustin, CA, 92780-7047

Description

Canon Aquilion Prime SP, Multislice Helical CT Scanner, TSX-303B/8L - Product Usage: intended to acquire and display cross-sectional volumes of the whole body, including the head.

Reason

A software problem has been identified which could result in the CT Scanner not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the CT scan failing resulting in rescanning and reinjection of contrast medium.

Action

On 11/9/2020, Canon issued a "Urgent: Medical Device Correction" letter to all affected customers via facsimile. In addition, to informing consignees about the recall, the firm ask affected consignees to take the following actions: 1. Until corrective action has been taken, in order to avoid the problem, please use the system with setting InstaView OFF according to the last notification (dated June 2020) Subject: Abnormality in Realtime Reconstruction Images. Please see Attachment A. 2. Please share this information with all users and reviewing radiologists as well as the clinical engineering or biomedical group at your facility. 3. If you have any questions, please feel free to contact Terry Schultz Regulatory Affairs Manager at (800) 421-1968 or your local Canon service representative at (800) 521-1968.

Distribution

US Nationwide distribution. OUS: Unknown

Quantity

60 units