FDA Recall Terminated

LIFEPAK EXPRESS defibrillator

Recall: Z-0548-06 · Initiated January 31, 2006

Recall

Recall Number
Z-0548-06
Event Number
34534
Firm
Medtronic Emergency Response Systems, Inc.
FEI Number
3015876
Product Code
MKJ
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
January 31, 2006
Posted
February 23, 2006
Terminated
January 17, 2011
Address
11811 Willows Rd Ne, Redmond, WA, 98052-2003

Description

LIFEPAK EXPRESS defibrillator

Reason

One of the printed circuit board assemblies may be contaminated with solder flux which can cause a short that may render the device inoperative.

Action

On 1/31/06 the firm sent a letter flagged URGENT-MEDICAL DEVICE RECALL to their customers via certified return receipt. The letter advised customers of the problem and requested the customers inspect their units at least weekly. The customers are provided additional guidance if they see certain symbols as described in the letter. The firm''s service personnel will visit the customers, remove and replace units.

Distribution

Units were distributed worldwide to public access environments.

Quantity

20,525 devices total - CR Plus and LIFEPAK Express