FDA Recall
Terminated
LIFEPAK EXPRESS defibrillator
Recall: Z-0548-06
·
Initiated January 31, 2006
Recall
- Recall Number
- Z-0548-06
- Event Number
- 34534
- Firm
- Medtronic Emergency Response Systems, Inc.
- FEI Number
- 3015876
- Product Code
- MKJ
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- January 31, 2006
- Posted
- February 23, 2006
- Terminated
- January 17, 2011
- Address
- 11811 Willows Rd Ne, Redmond, WA, 98052-2003
Description
LIFEPAK EXPRESS defibrillator
Reason
One of the printed circuit board assemblies may be contaminated with solder flux which can cause a short that may render the device inoperative.
Action
On 1/31/06 the firm sent a letter flagged URGENT-MEDICAL DEVICE RECALL to their customers via certified return receipt. The letter advised customers of the problem and requested the customers inspect their units at least weekly. The customers are provided additional guidance if they see certain symbols as described in the letter. The firm''s service personnel will visit the customers, remove and replace units.
Distribution
Units were distributed worldwide to public access environments.
Quantity
20,525 devices total - CR Plus and LIFEPAK Express