FDA Recall Terminated

FastPack Total PSA Immunoassay Chemilunescence assay for the determination of Total PSA. FastPack Total PSA Immunoassay is a paramagnetic particle immunoassay for the in vitro quantitative determination of prostate-specific antigen (PSA) in human serum and plasma as an aid in the management of patients with prostate cancer. The FastPack Total PSA Immunoassay is designed for use with the FastPack System.

Recall: Z-0547-2014 · Initiated November 8, 2013

Recall

Recall Number
Z-0547-2014
Event Number
66842
Firm
Qualigen Inc
FEI Number
3003355251
Product Code
LTJ
Status
Terminated
Root Cause
Process change control
Initiated
November 8, 2013
Posted
December 19, 2013
Terminated
September 18, 2014
Address
2042 Corte del Nogal, Carlsbad, CA, 92011-1438

Description

FastPack Total PSA Immunoassay Chemilunescence assay for the determination of Total PSA. FastPack Total PSA Immunoassay is a paramagnetic particle immunoassay for the in vitro quantitative determination of prostate-specific antigen (PSA) in human serum and plasma as an aid in the management of patients with prostate cancer. The FastPack Total PSA Immunoassay is designed for use with the FastPack System.

Reason

Qualigen initiated this recall because the FastPack kit may cause lower than expected results due to a manufacturing defect.

Action

Qualigen, Inc., notified customers of the recall by calling customers via telephone (with a telephone script) on 11/08/2013. The telephone (script) included the reason for recall, product description with codes, instructions and contact information. The customers were instructed to examine your inventory and discontinue use of the kits immediately. Qualigen will send you replacement kits and provide you with a shipping label to return all of the defective kits including any that were partially used. Feel free to contact System Support at 760.579.6900 if you have any questions or require assistance with returning your kits.

Distribution

US Distribution to states of: AL, CA, KS, KY, IN, MA, NC, NY, NV, OK, SC, TN, TX, UT, VA, and WA.

Quantity

182