FDA Recall Terminated

Brilliance 16 Slice (Power), Model 728240 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories

Recall: Z-0545-2019 · Initiated October 18, 2018

Recall

Recall Number
Z-0545-2019
Event Number
81569
Firm
Philips Medical Systems (Cleveland) Inc
FEI Number
1525965
Product Code
JAK
Status
Terminated
Root Cause
Process control
Initiated
October 18, 2018
Terminated
August 3, 2021
Address
595 Miner Rd, Cleveland, OH, 44143-2131

Description

Brilliance 16 Slice (Power), Model 728240 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories

Reason

There is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fall

Action

On October 18, 2018, Philips notices to their customers via courier service. Customers were advised to take the following actions: If any unusual noise is heard, discontinue use of the system immediately and contact your Philips representative for support. Field Order Change (FCO) 72800703 will be deployed to initiate an inspection of the front gantry cover. If the inspection finds evidence of weakening at this connection point, a temporary repair solution will be put in place to secure the cover for continued system use. FCO 72800704 will then be deployed to replace the affected front cover. A Philips Field Service Engineer will contact you to schedule the inspection of your system. Customers in need of further information or support concerning this issue, please contact your local Philips representative (North America & Canada) 1-800-722-9377.

Distribution

US Nationwide Distribution

Quantity

6,343 units in total