FDA Recall Open, Classified

Accessory kit, Alinity i containing Alinity ci Level Sensor, Bulk Solution

Recall: Z-0544-2020 · Initiated September 16, 2019

Recall

Recall Number
Z-0544-2020
Event Number
83847
Firm
Abbott Gmbh & Co. KG Max-Planck-Ring 2 65205 Wiesbaden Germany
FEI Number
3002809144
Product Code
JJE
Status
Open, Classified
Root Cause
Device Design
Initiated
September 16, 2019

Description

Accessory kit, Alinity i containing Alinity ci Level Sensor, Bulk Solution

Reason

Potential reliability issue with the Alinity ci series Level Sensor, Bulk Solution. The development of environmental stress on the bulk solution level sensor can cause cracks.

Action

Abbott GmbH & Co. KG notified customers on about 09/16/2019, via "Product Correction Urgent Immediate Action Required" letter. Customers were informed that the development of environmental stress on the bulk solution level sensor can cause cracks. Cracks can allow air to enter the fluid line, which may result either a failure to dispense or an incomplete dispense of the bulk solution on Alinity i and Alinity c. Instructions included that if using Alinity i, and you receive one of the following message codes: 1043, 1044, 1072, 1402, 1403. Refer to Appendix B in the letter. If using Alinity c, and you receive one of the following message codes: 3687, 3689. Refer to Appendix B in the letter. For both Alinity i and c, and those "if" scenarios occur, then discontinue reporting results until troubleshooting is complete and inspect the Alinity ciseries bulk solution level sensors per the instructions in Appendix A in the letter. If a bulk solution level sensor is found to be cracked: Replace the part before running the system using the replacement instructions found in the Alinity ciseries Operations Manual and Abbott recommends following your laboratory procedure to review results produced since your last successful Quality Control (QC). Last, if the level sensor is not cracked, refer to the Alinity ciseries Operations Manual to troubleshoot for the specific message code. To assist in the detection of a cracked bulk solution level sensor, Abbott recommends running Quality Control (QC) immediately before performing weekly maintenance. If your laboratory had previously identified cracks on the bulk solution level sensor before receipt of this communication, and there are messages associated with patient results in the Alinity ciseries System Logs during the same time period, Abbott recommends following your laboratory procedure to review results produced from the replacement of the bulk solution level sensor to the previous successful Quality Co

Distribution

Nationwide distribution to AL, AR, AZ, CA, CO, FL, GA, ID, IL, IN, KS, LA, MA, ME, MI, MN, MO, MS, NC, NJ, NV, NY, OK, PR, SC, SD, TN, TX, UT, VA, WA, WI. International distribution to ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVINA, BOTSWANA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, GUATEMALA, HONG KONG, INDIA, INDONESIA, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KUWAIT, LATVIA, LEBANON, LITHUANIA, MADAGASCAR, MALAYSIA, MEXICO, NAMIBIA, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PAKISTAN, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD & TOBAGO, TURKEY, UAE, UK, URUGUAY, VIETNAM.

Quantity

3,271 units