Laceration Tray, Catalog # 1892R1 Product Usage: Surgical procedure
Recall
- Recall Number
- Z-0544-2018
- Event Number
- 78950
- Firm
- Robert Busse & Co. Inc. d.b.a. Busse Hospital Disposables
- FEI Number
- 2433012
- Product Code
- PWY
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- June 27, 2017
- Terminated
- July 27, 2020
- Address
- 75 Arkay Dr, Hauppauge, NY, 11788-3707
Description
Laceration Tray, Catalog # 1892R1 Product Usage: Surgical procedure
BD SafetyGlide Needle 22GA 1-1/2 inch may contain the presence of loose polyprophylene foreign matter above release specification.
On June 27, 2017, Busse Hospital Disposables distributed Urgent Medical Device Recall notices and Response Forms to their customers via courier service. A copy of the supplier's recall notice was also provided to customers. Customers were advised to conduct the following: " Discontinue use of needle immediately. " Please check your inventory and cease distribution immediately. " If you have distributed the above products, please contact your customers and inform them of this recall immediately. " Contact Busse Hospital Disposables for a Return Good Authorization number. " Complete and Return the Response Form via fax to 631-435-2849 or via email to: [email protected]. Customers with questions may call 631-435-4711, ext. 254.
US Nationwide Distribution
75 cases