MultiDiagnost Eleva with FD. The Philips MultiDiagnost Eleva intended use is for the following applications: As a diagnostic-imaging device multi-functional/ universal system, general R/F, Fluoroscopy, Radiography and Angiography can be performed along with pediatric examinations and some more specialized interventional applications.
Recall
- Recall Number
- Z-0544-2011
- Event Number
- 57069
- Firm
- Philips Medical Systems North America Co. Phillips
- FEI Number
- 1000524572
- Product Code
- IZI
- Status
- Terminated
- Root Cause
- Other
- Initiated
- May 26, 2009
- Posted
- December 7, 2010
- Terminated
- December 9, 2010
- Address
- 22100 Bothell Everett Hwy, Bothell, WA, 98021-8431
Description
MultiDiagnost Eleva with FD. The Philips MultiDiagnost Eleva intended use is for the following applications: As a diagnostic-imaging device multi-functional/ universal system, general R/F, Fluoroscopy, Radiography and Angiography can be performed along with pediatric examinations and some more specialized interventional applications.
The Printed Circuit Board (PCB) in the Main-cabinet. Further investigation revealed that a manufacturing error in the PCB is responsible for the reliability issue. As a result, the complete system may shut down and not re-start.
Philips began mailing out an URGENT-Device Correction letter on May 26, 2009. Philips sent out a second URGENT-Device Correction letter dated June 4, 2009, to all consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were advised to contact their local Philips representative immediately if their system completely shut down and would not restart. A mandatory Field Change Order 72200144 was issued for replacement of the Printed Circuit Board (PCB) in the Allura XPER FD system. A mandatory Field Change Order FCO 70800091 was issued for replacement of the Printed Circuit Board (PCB) in the MultiDiagnost Eleva with FD system. The consignees were to be contacted by their Philips Representative for implementation of the corrective action. The firm's Field Engineer would replace the power tray of the systems. If the consignees needed any further information or support concerning this issue, they were to contact the call Center at 800-722-9377, # 5 and reference FCO 72200144 or FCO 70800091 with any questions.
Nationwide Distribution including the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MD, MI, MN, MO, MS, NC, NJ, NY, OH, OK, OR, PA, RI, TN, TX, VA, and WI.
5 units