FDA Recall Terminated

Varian brand Clinac, Accelerator, Linear, Medical charged-particle radiation therapy system: Model: High Energy Clinacs with one or more of the following options: -- BrainLAB micro MLC, -- 6MV SRS photon beam, -- Fine Beam Isocenter Accuracy, -- Stereotactic motion disable. ----Product is manufactured and distributed by Varian Medical Systems, Palo Alto, CA

Recall: Z-0544-2008 · Initiated September 10, 2007

Recall

Recall Number
Z-0544-2008
Event Number
45520
Firm
Varian Medical Systems Inc
FEI Number
2916710
Product Code
IYE
Status
Terminated
Root Cause
Labeling design
Initiated
September 10, 2007
Posted
February 14, 2008
Terminated
November 27, 2010
Address
911 Hansen Way, Palo Alto, CA, 94304

Description

Varian brand Clinac, Accelerator, Linear, Medical charged-particle radiation therapy system: Model: High Energy Clinacs with one or more of the following options: -- BrainLAB micro MLC, -- 6MV SRS photon beam, -- Fine Beam Isocenter Accuracy, -- Stereotactic motion disable. ----Product is manufactured and distributed by Varian Medical Systems, Palo Alto, CA

Reason

Treatment Error: The collimator drive chain may break or slip off of its drive track, allowing the collimator to rotate freely without motor control; if undetected resulting in a treatment with the wrong collimator angle.

Action

On 9/10/07, the firm issued an Urgent Medical Device Correction letter to all its direct consignees, informing them of the device defect and providing instructions on the recall. Short term user instructions were provided and the firm will schedule follow-up visits to make the needed permanent device corrections at each device site.

Distribution

Worldwide- Canada, Mexico, and the USA

Quantity

418 (total for Recalls Z-0544-0545-2008)