FDA Recall Open, Classified

Mammotrak Interventional Coil 1.5T

Recall: Z-0542-2025 · Initiated November 4, 2024

Recall

Recall Number
Z-0542-2025
Event Number
95645
Firm
Philips North America Llc
FEI Number
3016618143
Product Code
MOS
Status
Open, Classified
Root Cause
Use error
Initiated
November 4, 2024
Posted
November 25, 2024
Address
222 Jacobs St, Cambridge, MA, 02141-2289

Description

Mammotrak Interventional Coil 1.5T

Reason

Potential safety issue where a patient may be harmed while preparing for or during a scan.

Action

An URGENT MEDICAL DEVICE CORRECTION notice dated 10/31/24 was mailed to consignees. Consignees are to continue using affected systems in accordance with the intended use. When using the system follow section PRECAUTIONS, CAUTIONS & WARNINGS in the Instructions For Use (IFU). Consignees are to circulate the recall notice to all users for their awareness of the potential issue. The Recall Response Form is to be provided back to Philips MR within 30 days at [email protected]. Customers with any questions are to contact their local Philips Representative.

Distribution

US Nationwide distribution.

Quantity

5,231 units