FDA Recall
Terminated
The Is-Legionella IgG/IgM/IgA test kits are packaged in cardboard boxes with individual components secured within a foam insert.
Recall: Z-0541-05
·
Initiated December 23, 2004
Recall
- Recall Number
- Z-0541-05
- Event Number
- 31078
- Firm
- Diamedix Corporation
- FEI Number
- 1044713
- Product Code
- LHL
- Status
- Terminated
- Root Cause
- Other
- Initiated
- December 23, 2004
- Posted
- February 15, 2005
- Terminated
- May 3, 2012
- Address
- 2140 N Miami Ave, Miami, FL, 33127-4916
Description
The Is-Legionella IgG/IgM/IgA test kits are packaged in cardboard boxes with individual components secured within a foam insert.
Reason
The kits belonging to lot #s 30804/30804Z and 91803Z may lose activity, after one or two usages, causing the OD values for the Cut-Off Calibrator and/or Positive Control to be recovered below their specified ranges.
Action
Direct customers were notified by letter on 12/23/2004 by Federal Express.100% of the total customers will be contacted. The letter instructed the consignees to discontinue use immediately and arrange to return the product to Diamedix.
Distribution
Distributed to Domestic customers only.
Quantity
178