FDA Recall Terminated

Discovery, Slide Staining System, Catalog Number: N750-DIS-FS, Ventana Medical Systems Inc, Tucson, AZ

Recall: Z-0540-2008 · Initiated September 29, 2006

Recall

Recall Number
Z-0540-2008
Event Number
45496
Firm
Ventana Medical Systems Inc
FEI Number
1000125183
Product Code
KPA
Status
Terminated
Root Cause
Device Design
Initiated
September 29, 2006
Posted
January 24, 2008
Terminated
June 6, 2013
Address
1910 E Innovation Park Dr, Tucson, AZ, 85755-1962

Description

Discovery, Slide Staining System, Catalog Number: N750-DIS-FS, Ventana Medical Systems Inc, Tucson, AZ

Reason

Leaking Carboys/Safety Updates:The Ventana Staining Platform includes one or more carboys that may have the potential to leak if the spigot is either improperly installed or not fully closed.

Action

The firm has mailed three customer letters, a) the first letter dated 9/30/2005 describes the upgraded Voila CD that contains safety information; b) a second letter, dated 6/27/2006, identified as a "Customer Alert" was mailed describing an upgrade to the software system version 9.3/10.3 and the CD that includes safety improvements, c) the third letter dated 9/29/2006 and identified as a Safety Alert advises that as part of their effort to continuously improve laboratory safety, they are updating the labeling and instructions for use regarding installation and cleaning for the carboy as the Ventana staining platform includes one or more carboys that may have the potential to leak if the spigot is either improperly installed or not fully closed. Enclosed are pictogram labels that indicate "slip hazard" and "refer to instructions for use." The customer is instructed to apply these two labels to the carboys above the spigot as a reminder to laboratory personnel to keep the spigot in the full "off" position when not in use and to refer to the instructions for use for maintenance. In addition, the letter advises that there is an electrical shock warning for the power strip that is part of their system to indicate that it should be mounted off of the floor. This power strip is intended to connect the computer components of the staining system and should not be placed on the floor where fluid might contact it. Also included is a warning label for shock hazard to apply to their power strip.

Distribution

Worldwide; USA, Australia, China, Canada, Germany, Great Britain, France, Italy, Hungary, Singapore, Korea, Taiwan, Spain, Poland, Greece, Austria, Sweden, Japan, Israel, Denmark, Turkey, Finland, Saudi Arabia, The Netherlands, Norway, Slovenia, Ireland, and Belgium

Quantity

USA - 1,712 units; Foreign - 1,159 units (total for recalls-Z-0537-0541-2008)