FDA Recall Terminated

Giraffe OmniBed infant warmer

Recall: Z-0538-2020 · Initiated October 11, 2019

Recall

Recall Number
Z-0538-2020
Event Number
84077
Firm
GE Healthcare, LLC
FEI Number
2126677
Product Code
FMT
Status
Terminated
Root Cause
Device Design
Initiated
October 11, 2019
Posted
December 3, 2019
Terminated
August 15, 2024
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

Giraffe OmniBed infant warmer

Reason

Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall.

Action

The firm disseminated its notices by letter beginning on 10/11/2019. The notices identified the following safety issues: - The bedside panels of Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation can be upright and look closed but not be latched. - The portholes also can look closed when not latched. - If a canopy cover is used, it can hold the bedside panel or porthole door closed without being latched. The product may continue to be used following specific safety instructions outlined in the notice. On/about 11/8/2019, the firm issued an "Amended - Urgent Medical Device Recall" letter that supplemented the previous notification to provide additional safety instructions and materials (poster regarding risk of patient fall, user manual addendum, and labeling, including safety labels). Update 6/3/2021: In the event a new consignee was identified, GE Healthcare has sent the original letter.

Distribution

worldwide

Quantity

23,693 units