FDA Recall
Terminated
Welch Allyn Acuity Central Monitoring Station (With SunBlade Model 150,650 MHZ. Part # 700-0362-00
Recall: Z-0538-04
·
Initiated December 31, 2003
Recall
- Recall Number
- Z-0538-04
- Event Number
- 28003
- Firm
- Welch Allyn Protocol, Inc
- FEI Number
- 3023750
- Product Code
- DSI
- Status
- Terminated
- Root Cause
- Other
- Initiated
- December 31, 2003
- Posted
- February 13, 2004
- Terminated
- January 6, 2005
- Address
- 8500 SW Creekside Pl, Beaverton, OR, 97008-7107
Description
Welch Allyn Acuity Central Monitoring Station (With SunBlade Model 150,650 MHZ. Part # 700-0362-00
Reason
Potential failure to audibly alarm due to speaker hardware component failure.
Action
The firm sent a letter dated December 31, 2003 instructing their customers to install an audio cable from the CPU to the Flat Panel Monitor to provide redundant speaker capability. The audio cable and installation instructions were included in the letter. The letter also included a Correction Response Form which customers should fill out upon completion of correction.
Distribution
The firm distributes the product nationwide and internationally.
Quantity
233 devices