FDA Recall
Open, Classified
dS Breast 16ch 1.5T
Recall: Z-0537-2025
·
Initiated November 4, 2024
Recall
- Recall Number
- Z-0537-2025
- Event Number
- 95645
- Firm
- Philips North America Llc
- FEI Number
- 3016618143
- Product Code
- MOS
- Status
- Open, Classified
- Root Cause
- Use error
- Initiated
- November 4, 2024
- Posted
- November 25, 2024
- Address
- 222 Jacobs St, Cambridge, MA, 02141-2289
Description
dS Breast 16ch 1.5T
Reason
Potential safety issue where a patient may be harmed while preparing for or during a scan.
Action
An URGENT MEDICAL DEVICE CORRECTION notice dated 10/31/24 was mailed to consignees. Consignees are to continue using affected systems in accordance with the intended use. When using the system follow section PRECAUTIONS, CAUTIONS & WARNINGS in the Instructions For Use (IFU). Consignees are to circulate the recall notice to all users for their awareness of the potential issue. The Recall Response Form is to be provided back to Philips MR within 30 days at [email protected]. Customers with any questions are to contact their local Philips Representative.
Distribution
US Nationwide distribution.
Quantity
5,231 units