FDA Recall
Terminated
Pump Tube, REF 8170,121; Suction Pump Drain Tube; An accessory to the suction pump designed for use with the Richard Wolf Ultrasound Lithotriptor used to disintegrate kidney stones, urinary bladder stones and ureter stones; Richard Wolf GmbH, D-75438 Knittlingen Germany
Recall: Z-0537-04
·
Initiated February 2, 2004
Recall
- Recall Number
- Z-0537-04
- Event Number
- 28227
- Firm
- Richard Wolf Medical Instruments Corp.
- FEI Number
- 1418479
- Product Code
- JOL
- Status
- Terminated
- Root Cause
- Other
- Initiated
- February 2, 2004
- Posted
- February 11, 2004
- Terminated
- March 24, 2004
- Address
- 353 Corporate Woods Pkwy, Vernon Hills, IL, 60061-3110
Description
Pump Tube, REF 8170,121; Suction Pump Drain Tube; An accessory to the suction pump designed for use with the Richard Wolf Ultrasound Lithotriptor used to disintegrate kidney stones, urinary bladder stones and ureter stones; Richard Wolf GmbH, D-75438 Knittlingen Germany
Reason
If the drain tube is assembled incorrectly, the suction pump will not operate as intended.
Action
The accounts were telephoned on 2/2/04 and informed of the potential to re-assemble the suction pump drain tubing incorrectly after sterilization. The accounts were sent re-designed replacement tubing sets and were requested to return the old tubing sets to Richard Wolf.
Distribution
Texas and Nebraska
Quantity
2 tube sets