FDA Recall
Terminated
Biomet Polaris 5.5 Button Lock Screw Inserter. Catalog number 14-500178
Recall: Z-0536-2013
·
Initiated September 29, 2012
Recall
- Recall Number
- Z-0536-2013
- Event Number
- 63653
- Firm
- Ebi, Llc
- FEI Number
- 2242816
- Product Code
- HXX
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- September 29, 2012
- Posted
- December 17, 2012
- Terminated
- December 7, 2015
- Address
- 100 Interpace Pkwy, Parsippany, NJ, 07054-1149
Description
Biomet Polaris 5.5 Button Lock Screw Inserter. Catalog number 14-500178
Reason
Broken screw insertor hardness was measured and found to be below the acceptable limits.
Action
Biomet mailed Urgent Medical Device Recall Notice/Fax Back Response Form dated 9/25/2012 to the distributors on 9/29/2012.
Distribution
Nationwide Distribution, including the states of MO, TX, and IN.
Quantity
15 units