Ambu, Positive End Expiratory Pressure (PEEP) Valves, PEEP Valve Spare Part, REF 137 000 501, Manufacturer: Ambu International A/S, DK2750 Ballerup USA: Ambu Inc. (Ambu USA) Linthicum, MD 21090-1356 USA, Made in Denmark
Recall
- Recall Number
- Z-0536-2008
- Event Number
- 44734
- Firm
- Ambu Incorporated
- FEI Number
- 1121788
- Product Code
- BYE
- Status
- Terminated
- Root Cause
- Process design
- Initiated
- August 21, 2007
- Posted
- December 22, 2007
- Terminated
- June 16, 2008
- Address
- 6740 Baymeadow Drive, Glen Burnie, MD, 21060-6412
Description
Ambu, Positive End Expiratory Pressure (PEEP) Valves, PEEP Valve Spare Part, REF 137 000 501, Manufacturer: Ambu International A/S, DK2750 Ballerup USA: Ambu Inc. (Ambu USA) Linthicum, MD 21090-1356 USA, Made in Denmark
Lower pressure than expected: Medical device for respiratory care may leak and not register accurate flow settings that affect patient respiration.
Ambu notified consignees with a letter dated 08/21/07 and flagged as Trouble Shooting Guide. The letter advised that some valves do not hold the set PEEP level under certain conditions and consignees were further advised to consider the guideline as an addendum to existing directions for use. Consignees were instructed to return (to Ambu) for replacement the identified lots listed in the notification and confirm receipt of the advisory by e-m or fax. Ambu subsequently identified additional lots of valves and replacement valve parts for recall and notified all consignees by "Urgent Device Recall" letter dated 11/29/07. The letter advised healthcare practitioners of the impact on patient treatment that may result in less than effective treatment. Consignees were further advised to cease use of the device and return for replacement.
Nationwide.
197 total