FDA Recall Terminated

Depuy DELTA Xtend, Standard Humeral PE Cup, Dia 38/+6R RET, sterile, Depuy France; REF 130738106. The device is used as a shoulder implant.

Recall: Z-0535-2009 · Initiated July 10, 2008

Recall

Recall Number
Z-0535-2009
Event Number
49112
Firm
Depuy Orthopaedics, Inc.
FEI Number
1818910
Product Code
KWS
Status
Terminated
Root Cause
Device Design
Initiated
July 10, 2008
Posted
January 15, 2009
Terminated
October 28, 2009
Address
700 Orthopaedic Dr, Warsaw, IN, 46582-3994

Description

Depuy DELTA Xtend, Standard Humeral PE Cup, Dia 38/+6R RET, sterile, Depuy France; REF 130738106. The device is used as a shoulder implant.

Reason

Compressed air may become trapped between the cup and the stem, preventing the tapered walls from engaging and causing the cup to feel unstable.

Action

A letter, "Urgent Device Correction" dated July 8, 2008 was sent via certified mail to implanting surgeons advising them of the issues and guidance on how to mitigate the problem. A communication was also sent to all sales representatives and territory offices on July 15, 2008. For questions, please contact Depuy Orthopaedics, Inc. Customer Service at 1-800-473-3789.

Distribution

Nationwide Distribution.

Quantity

1,521 of all devices.