Alaris PC unit model 8000 and 8015 with Alaris PCA module model 8120 and Auto ID barcode scanner model 8600. Product Usage: The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System. The Alaris PCA (Patient Controlled Analgesia) pump is one of the modules offer under the Alaris System. The Auto ID is a barcode scanner.
Recall
- Recall Number
- Z-0534-2014
- Event Number
- 66865
- Firm
- CareFusion 303, Inc.
- FEI Number
- 2016493
- Product Code
- FRN
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- December 5, 2013
- Posted
- December 20, 2013
- Terminated
- June 2, 2016
- Address
- 10020 Pacific Mesa Blvd, San Diego, CA, 92121-4386
Description
Alaris PC unit model 8000 and 8015 with Alaris PCA module model 8120 and Auto ID barcode scanner model 8600. Product Usage: The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System. The Alaris PCA (Patient Controlled Analgesia) pump is one of the modules offer under the Alaris System. The Auto ID is a barcode scanner.
CareFusion is recalling the Alaris PC unit (model 8000 and 8015) because the Alaris PC unit will produce an audible alarm and the Alaris PC unit will display a System Error message.
CareFusion sent an Urgent Medical Device Recall Notification letter dated December 5, 2013, to all affected customers. The letter identifed the affected product, problem and actions to be taken. Customers were instructed to contact CareFusion Customer Advocacy at (888) 812-3266, if they experienced a SYSTEM ERROR on the Alaris PC unit. CareFusion will contact all affected customers within 60 days to initiate the scheduling process for the remediation. Customers are instructed to complete and return the enclosed Customer Response Card to expedite the corrective action process. Customers with questions were instructed to contact CareFusion Support Center at 888-562-6018, or for Technical Support call 888-812-3229.
Worldwide Distribution - USA (nationwide) and Internationally to Canada, Netherlands, and Australia.
4824 total units (4242 units in US)