FDA Recall Terminated

Ambu, Positive End Expiratory Pressure (PEEP) Valves, PEEP Valve 10, 22/18 mm, REF 000 138 000, Manufacturer: Ambu International A/S, DK2750 Ballerup USA: Ambu Inc. (Ambu USA) Linthicum, MD 21090-1356 USA, Made in Denmark

Recall: Z-0534-2008 · Initiated August 21, 2007

Recall

Recall Number
Z-0534-2008
Event Number
44734
Firm
Ambu Incorporated
FEI Number
1121788
Product Code
BYE
Status
Terminated
Root Cause
Process design
Initiated
August 21, 2007
Posted
December 22, 2007
Terminated
June 16, 2008
Address
6740 Baymeadow Drive, Glen Burnie, MD, 21060-6412

Description

Ambu, Positive End Expiratory Pressure (PEEP) Valves, PEEP Valve 10, 22/18 mm, REF 000 138 000, Manufacturer: Ambu International A/S, DK2750 Ballerup USA: Ambu Inc. (Ambu USA) Linthicum, MD 21090-1356 USA, Made in Denmark

Reason

Lower pressure than expected: Medical device for respiratory care may leak and not register accurate flow settings that affect patient respiration.

Action

Ambu notified consignees with a letter dated 08/21/07 and flagged as Trouble Shooting Guide. The letter advised that some valves do not hold the set PEEP level under certain conditions and consignees were further advised to consider the guideline as an addendum to existing directions for use. Consignees were instructed to return (to Ambu) for replacement the identified lots listed in the notification and confirm receipt of the advisory by e-m or fax. Ambu subsequently identified additional lots of valves and replacement valve parts for recall and notified all consignees by "Urgent Device Recall" letter dated 11/29/07. The letter advised healthcare practitioners of the impact on patient treatment that may result in less than effective treatment. Consignees were further advised to cease use of the device and return for replacement.

Distribution

Nationwide.

Quantity

197 total