FDA Recall
Terminated
Toshiba America Medical Systems, Ultimax System MDX-8000A equipped with the EPS-Plus/ADR-1000A/E2D Digital Spot Film Device
Recall: Z-0533-06
·
Initiated February 23, 2006
Recall
- Recall Number
- Z-0533-06
- Event Number
- 34955
- Firm
- Toshiba American Med Sys Inc
- FEI Number
- 2020563
- Product Code
- LLZ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- February 23, 2006
- Posted
- March 29, 2006
- Terminated
- February 29, 2012
- Address
- 2441 Michelle Dr, Tustin, CA, 92781
Description
Toshiba America Medical Systems, Ultimax System MDX-8000A equipped with the EPS-Plus/ADR-1000A/E2D Digital Spot Film Device
Reason
When the error condition occurs the x-ray exposures will terminate immediately. The digital system (EPS-Plus/ADR-1000A) will immediately stop acquiring images or will continue to acquire blank images. The remote console will lock-up upon release of the hand switch. The operator must reboot the system. This will result in the loss of the images acquired during the contrast portion of the study.
Action
Recall letter to be sent to to customers of affected devices beginning February 23, 2006. Corrected software to be issued by Toshiba America Medical Systems Inc, for all affected devices.
Distribution
Nationwide to AZ, CA, FL, MT, & PA
Quantity
5