FDA Recall Terminated

Advisor Vital Signs Monitor model 9200, catalog number 92E654220. item Description 9200 II 3LD/RSP/SP02 MDD MEDEXL. The basic monitoring package includes ECG (3 lead / 5 lead), impedance respiration, non-invasive blood pressure, pulse oximetry, two invasive blood pressures, and tow temperature channels. The device permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals.

Recall: Z-0532-05 · Initiated January 5, 2005

Recall

Recall Number
Z-0532-05
Event Number
30811
Firm
Smiths Medical PM, Inc.
FEI Number
2183502
Product Code
DPS
Status
Terminated
Root Cause
Other
Initiated
January 5, 2005
Posted
February 10, 2005
Terminated
November 14, 2005
Address
N7w22025 Johnson Dr, Waukesha, WI, 53186-1856

Description

Advisor Vital Signs Monitor model 9200, catalog number 92E654220. item Description 9200 II 3LD/RSP/SP02 MDD MEDEXL. The basic monitoring package includes ECG (3 lead / 5 lead), impedance respiration, non-invasive blood pressure, pulse oximetry, two invasive blood pressures, and tow temperature channels. The device permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals.

Reason

An issue with the pace detect function was discovered on a 9200 II monitor. Could result in a falsely high or low heart rate reading.

Action

On 12/22/04 distribution of new units was placed on hold and customers were notified to recommend that they do not use the pace detect function. On 12/27/03 an email was sent to customers of the solution. On 12/29/04 a Technical Worksheet describing how to replace a resistor at R8 was emailed to customers. Between 12/29/04 and 01/03/05 kits containing instructions and new resistors were sent to customers to begin reworking of the devices.

Distribution

Internationally only.

Quantity

2