FDA Recall Terminated

BladderScan Prime Plus Probe REF 0570-0395 - Product Usage: intended to be used for measuring the urine volume in the bladder noninvasively.

Recall: Z-0531-2021 · Initiated September 29, 2020

Recall

Recall Number
Z-0531-2021
Event Number
86528
Firm
Verathon, Inc.
FEI Number
3022472
Product Code
ITX
Status
Terminated
Root Cause
Vendor change control
Initiated
September 29, 2020
Terminated
June 29, 2022
Address
20001 N Creek Pkwy, Bothell, WA, 98011-8218

Description

BladderScan Prime Plus Probe REF 0570-0395 - Product Usage: intended to be used for measuring the urine volume in the bladder noninvasively.

Reason

Due to a change in handle material and process, probe handle may crack.

Action

On 10/5/2020, the firm sent a "Field Defect Correction URGENT MEDICAL DEVICE RECALL" Notification to its U.S, Customers via FedEx informing them that due to an unauthorized manufacturing change, the probe handle's plastic may become brittle and crack. Customer are instructed to: 1. Review the information contained within this Recall Notice with all staff members who should be aware of the contents of the letter. 2. Check the list of Serial Numbers that are provided within the Recall Response Form. 3. Perform a visual inspection of the probe handle for cracks or damage. 4. Based on the visual inspection of your device(s), please do one of the following: a. Unboxing and Initial Use " If your BladderScan Prime Plus probe handle plastics have no visible cracks or damage and / or you do not wish for units to be serviced, complete the attached Recall Response Form by indicating no or 0 in the text box and return the form to Verathon by email to [email protected]. Or " If your BladderScan Prime Plus probe handle plastics have visible cracks and you wish for these units to be serviced, complete attached Recall Response Form by indicating by number the count of probe handle plastics that require repair in the text box and return the form by email to [email protected]. Verathon Customer Care will contact you to arrange for repair of the probe handle plastics. This replacement will be at no charge to you. b. Subsequent Use If subsequent to submitting your Recall Response Form your BladderScan Prime Plus probes develop cracks (reference Figure 1) during subsequent use and you wish for them to be serviced, please email [email protected]. Verathon Customer Care will contact you to arrange for repair of the probe handle plastics. This replacement will be at no charge to you. Follow the instructions within the BladderScan Prime Plus Operations and Maintenance Manual (0900-4510 Rev. 11), and as stated below, for conduct

Distribution

Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, NC, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WI, WV and WY. The countries of Austrailia, Belgium, Canada, France, Germany, Kuwait, Luxembourg, New Zealand, Portugal, Singapore, Sweden, Switzerland, Taiwan, and United Kingdom.

Quantity

398 units