FDA Recall Terminated

Abbott Healon EndoCoat Ophthalmic Viscosurgical Device 3% sodium hyaluronate 0.85 mL 30 mg/mL Sterile A Solution Sterile EO Packaging and Cannula Product of USA Abbott Healon Duet Dual Pack 0.55 ml OVD 0.85 ml Healon EndoCoat OVD Contents: -one sterile syringe of Healon Ophthalmic Viscosurgical Device (OVD) 0.55 ml (10 mg/ml Sodium Hyaluronate) -one sterile syringe of Healon EndoCoat Ophthalmic Viscosurgical Device (OVD) 0.85 ml (30 mg/ml Sodium Hyaluronate) Part Number: US: VT585U and 57502000 (packaged in Healon Duet 10290080) OUS: VT585 and 57550500 (packaged in Healon duet 10220010, 10220011 and 10220012)

Recall: Z-0530-2015 · Initiated November 6, 2014

Recall

Recall Number
Z-0530-2015
Event Number
69704
Firm
Abbott Medical Optics Inc (AMO)
FEI Number
3003843509
Product Code
LZP
Status
Terminated
Root Cause
Unknown/Undetermined by firm
Initiated
November 6, 2014
Posted
December 12, 2014
Terminated
May 19, 2015
Address
1700 E Saint Andrew Pl, Santa Ana, CA, 92705-4933

Description

Abbott Healon EndoCoat Ophthalmic Viscosurgical Device 3% sodium hyaluronate 0.85 mL 30 mg/mL Sterile A Solution Sterile EO Packaging and Cannula Product of USA Abbott Healon Duet Dual Pack 0.55 ml OVD 0.85 ml Healon EndoCoat OVD Contents: -one sterile syringe of Healon Ophthalmic Viscosurgical Device (OVD) 0.55 ml (10 mg/ml Sodium Hyaluronate) -one sterile syringe of Healon EndoCoat Ophthalmic Viscosurgical Device (OVD) 0.85 ml (30 mg/ml Sodium Hyaluronate) Part Number: US: VT585U and 57502000 (packaged in Healon Duet 10290080) OUS: VT585 and 57550500 (packaged in Healon duet 10220010, 10220011 and 10220012)

Reason

AMO has received complaints where the finger grip dislodged while depressing the syringe plunger. The finger grip dislodgement could result in the surgeon's hand unintentionally moving the cannula further than intended, resulting in potential trauma to the eye.

Action

Abbott Medical Optics issued an Urgent Advisory Notice dated November 4, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised to consult the DFU that is distributed with each device, which contains important warning and precautions for use of the product. Customers were instructed to provide a copy of the Advisory Notice to surgeons and thsoe who need to be aware in their organization. Custommers with questions were instructed to call 877-266-4543. For questions regarding this recall call 714-247-8656.

Distribution

Worldwide Distribution - USA (Nationwide and Puerto Rico) and internationally to the following countries; Italy, Austria, Saudi Arabia, Turkey, Hong Kong, Canada, Germany, Australia, Columbia, Venezuela, Netherlands, Portugal, New Zealand, Denmark, Argentina, France, Belgium, Costa Rica, Singapore, Ukraine, Guadeloupe, Kuwait, Chile, Peru, Mexico, Czech Republic, Serbia, Hungary, Slovenia, Guatemala, Finland, Honduras, Malaysia, United Kingdom, Spain, Switzerland, Israel, and Croatia.

Quantity

484,881 Nationwide (U.S. and Puerto Rico) and 483,984 Internationally