FDA Recall Open, Classified

MAGEC 1 System

Recall: Z-0528-2022 · Initiated June 25, 2019

Recall

Recall Number
Z-0528-2022
Event Number
89177
Firm
NuVasive Specialized Orthopedics, Inc.
FEI Number
3006179046
Product Code
PGN
Status
Open, Classified
Root Cause
Device Design
Initiated
June 25, 2019
Address
101 Enterprise, Ste 100, Aliso Viejo, CA, 92656-2604

Description

MAGEC 1 System

Reason

Safety Alert to all surgeons informing them of post-implantation fractures of an internal metallic component for device manufactured prior to March 26, 2015.

Action

Beginning on 06/25/2019, a Field Safety Notice was communicated to all surgeon informing them that post-implantation fracture of an internal metallic component (i.e., locking pin) has been observed in early versions of the MAGEC System Rods that were manufactured prior to March 26, 2015. Fracture of the locking pin may affect the ability of the device to lengthen and may be associated with Titanium wear debris generation and localized tissue discoloration. Recommended User Actions: - NuVasive, Inc. voluntarily issues this Safety Alert to reinforce the Instructions for Use (IFU) of the MAGEC System to clinical users - The IFU should be consulted on an ongoing basis throughout patient treatment with the MAGEC System. - Users should follow the appropriate postoperative procedure to assess the MAGEC System by X-ray imaging whenever the device is adjusted or at a minimum of once every six months. Typically, the device can be adequately visualized, and a fractured locking pin can be detected, on standard anteroposterior X-ray imaging. Figure 1 shows representative X-ray images for comparison of a device with an intact locking pin (A) to a device manufactured prior to March 26, 2015 with a fractured locking pin (B). A fractured locking pin can typically be identified by examining the noted areas to detect the separation of the internal components and/or the extension of an internal component beyond the device housing. - If a fractured locking pin is detected, removal of the device may be indicated. The decision to remove the device should be made by the physician in consultation with the patient and/or family. - If removal is deemed appropriate by the physician, the device and all associated accessories should be removed and the explanted device should be returned to NuVasive. - Clinical users should review the IFU and be aware of the warnings and precautions on an ongoing basis throughout patient treatment with the device. - Patients and/or families should be reminde

Distribution

Worldwide

Quantity

3,056 devices