FDA Recall Terminated

Broncho Flexible Fiberscope, models: 11001BN1, 11002BD1, 11004BC1, and 11009BC1. " Examination and treatment of the tracheobronchial tree Contraindications for Bronchoscopy " Uncontrolled angina, uncontrolled heart failure, or serious uncontrolled ventricular arrhythmias " Uncontrolled hypertension " Bleeding disorder or use of anticoagulants.

Recall: Z-0527-2014 · Initiated November 13, 2013

Recall

Recall Number
Z-0527-2014
Event Number
66862
Firm
Karl Storz Endoscopy America Inc
FEI Number
2020550
Product Code
ECQ
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
November 13, 2013
Posted
December 17, 2013
Terminated
March 27, 2014
Address
15 Wells St, Southbridge, MA, 01550-4503

Description

Broncho Flexible Fiberscope, models: 11001BN1, 11002BD1, 11004BC1, and 11009BC1. " Examination and treatment of the tracheobronchial tree Contraindications for Bronchoscopy " Uncontrolled angina, uncontrolled heart failure, or serious uncontrolled ventricular arrhythmias " Uncontrolled hypertension " Bleeding disorder or use of anticoagulants.

Reason

KARL STORZ Endoscopy America, Inc. is recalling the Flexible Broncho Fiberscope or a Flexible Intubation Fiberscope with a suction port because the material used in the suction channel portion was thermoplastic polyurethane (TPU), which is not listed in STERRAD systems claims for interior endoscope lumens.

Action

The firm, KARL STORZ, sent a "A SAFETY ALERT-ATTENTION NEEDED" letter dated November 8, 2013 to all their customers on November 13, 2013. The letter describes the product, problem and actions to be taken.The letter informs the customers that the endoscopes incorporated a lumen material that is only approved for use as an external surface material and not included in the STERRAD NX and 100NX FDA-cleared claims for use in flexible endoscope lumens. The customers were instructed to follow the recommended processing methods included in the letter. Customers with questions about the letter are instructed to contact the undersigned and the Technical Services at (800) 421-0837, ext 5350.

Distribution

Nationwide Distribution.

Quantity

141 units