CDLCAM, First-Pass Radionuclide Angiography Camera System, Model 70-500101, with operating Software Version B1A, marketed by CDL Medical Tech.
Recall: Z-0527-04
·
Initiated January 16, 2004
Recall
- Recall Number
- Z-0527-04
- Event Number
- 28161
- Firm
- GVI Technology Partners
- FEI Number
- 3003917438
- Product Code
- IYX
- Status
- Terminated
- Root Cause
- Other
- Initiated
- January 16, 2004
- Posted
- February 10, 2004
- Terminated
- April 12, 2012
- Address
- 1470 Enterprise Pkwy, Twinsburg, OH, 44087-2242
Description
CDLCAM, First-Pass Radionuclide Angiography Camera System, Model 70-500101, with operating Software Version B1A, marketed by CDL Medical Tech.
Reason
Due to limitations in the operating software, the acquired scan may not be processed properly.
Action
The firm notified their distributor by letter on 1/16/2004, software upgrade will be installed on each unit.
Distribution
One Single customer.
Quantity
16 units