FDA Recall Open, Classified

1. Footrest ( use exclusively in combination with Intelli-C devices) 2. Shoulder support ( use exclusively in combination with Intelli-C devices)

Recall: Z-0526-2022 · Initiated January 21, 2022

Recall

Recall Number
Z-0526-2022
Event Number
89413
Firm
NRT X-RAY A/S Birkegardsvej 16 BIRKEGAARDSVEJ 16 Hasselager Denmark
FEI Number
1000188474
Product Code
OWB
Status
Open, Classified
Root Cause
Process control
Initiated
January 21, 2022

Description

1. Footrest ( use exclusively in combination with Intelli-C devices) 2. Shoulder support ( use exclusively in combination with Intelli-C devices)

Reason

Friction rubber block in the clamping mechanism was missing in the Footrest. It may slip down along the patient table while in use with a patient. This accessory is designed to use with Intelli-C, Right and Intelli-C, Left.

Action

On 01/11/2022, NRT issued an Urgent Medical device correction, which includes the following action to be taken by the users. 1. Identified the affected accessories as described in the Urgent Medical Device Correction Letter. 2. Inspect the Footrest and Shoulder Support which use the same clamping mechanism and should be inspected in the same way. The clamping mechanism consists of two clamping profiles, one for each side of the tabletop. Both sides should be inspected. a. The friction rubber that must be inspected is placed inside the clamping profile. It is the thick rubber, which must be firmly in place. The rubber is glued on the profile, and you should look for the overall adhesion. If slightly pressed from each side and end, the rubber should not show signs of detachment from the profile. If the rubber is firmly in place, you may continue use the equipment. If the rubber is detaching along either side or end, do not use the accessory until the repair has been made. 3. Return the response card in page 4, with the results and date of the inspection and your signature, not later than February 15, 2022. to [email protected]. 4. Follow up will be made in regards to the repair in April 2022, which will be coordinated with US dealer Alpha Imaging in Willoughby, Ohio. For any question contact (300) 696-0004 or E-Mail at [email protected].

Distribution

USA: IA, KY,LA, MA, NC,NY, OH, SC, WA, WV OUS: Switzerland, Czech Republic, Germany, Denmark, Netherlands, RSA, Sweden

Quantity

72 units