FDA Recall
Terminated
Philips BV300 Mobile X-Ray System
Recall: Z-0526-03
·
Initiated September 27, 2002
Recall
- Recall Number
- Z-0526-03
- Event Number
- 25248
- Firm
- Philips Medical Systems
- FEI Number
- 1000524572
- Product Code
- IZL
- Status
- Terminated
- Root Cause
- Other
- Initiated
- September 27, 2002
- Posted
- February 1, 2003
- Terminated
- January 12, 2005
- Address
- 22100 Bothell Everett Highway, Bothell, WA, 98041
Description
Philips BV300 Mobile X-Ray System
Reason
The C-bow support can bend and break allowing C-arm to fall from support
Action
The firm issued a letter dated 9/27/02 to their customers. The letter advised of the problem and that a service representative will visit to inspect unit and correct.
Distribution
The firm distributed to hospitals, x-ray centers and medical centers throughout the U.S.
Quantity
229 devices